Download Catalog
Flame Photometer LBN-FP170 performs chemical analysis of alkali metals depending on their emission spectra. Calibrated device determines concentration of metals in the sample efficiently. Designed with 7-inch colour touch screen to display concentration values conveniently. Enables selection of different concentration units and allows pre- selection of flame sizes.
| Testing elements | K, Na, Li, Ca, Ba |
|---|---|
| No of channels | 5 |
| Display Value | Concentration value |
| Data range | 0.000 to 9.999 |
| Range | <table cellspacing=0 border=1> <tr> <td style=min-width:50px>K</td> <td style=min-width:50px>0 to 100 ppm</td> </tr> <tr> <td style=min-width:50px>Na</td> <td style=min-width:50px>0 to 160 ppm</td> </tr> <tr> <td style=min-width:50px>Li</td> <td style=min-width:50px>0 to 100 ppm</td> </tr> <tr> <td style=min-width:50px>Ca</td> <td style=min-width:50px>0 to 1000 ppm</td> </tr> <tr> <td style=min-width:50px>Ba</td> <td style=min-width:50px>0 to 3000 ppm</td> </tr> </table> |
| LOD | <table cellspacing=0 border=1> <tr> <td style=min-width:50px>K</td> <td style=min-width:50px>0.01 ppm</td> </tr> <tr> <td style=min-width:50px>Na</td> <td style=min-width:50px>0.01 ppm</td> </tr> <tr> <td style=min-width:50px>Li</td> <td style=min-width:50px>0.1 ppm</td> </tr> <tr> <td style=min-width:50px>Ca</td> <td style=min-width:50px>2 ppm</td> </tr> <tr> <td style=min-width:50px>Ba</td> <td style=min-width:50px>6 ppm</td> </tr> </table> |
| Linear error | <table cellspacing=0 border=1> <tr> <td style=min-width:50px>K</td> <td style=min-width:50px>0.195 ppm</td> </tr> <tr> <td style=min-width:50px>Na</td> <td style=min-width:50px>0.69 ppm</td> </tr> <tr> <td style=min-width:50px>Li</td> <td style=min-width:50px>0.15 ppm</td> </tr> <tr> <td style=min-width:50px>Ca</td> <td style=min-width:50px>3 ppm</td> </tr> <tr> <td style=min-width:50px>Ba</td> <td style=min-width:50px>9 ppm</td> </tr> </table> |
| Display | 7-inch colour touch screen |
| Response time | < 8 s |
| Sample uptake | <6 L/min |
| Stability | <3% drift over 15s when continuously aspirating |
| Reproducibility | <3% coefficient of variation for 7 consecutive samples |
| Fuel | LPG |
| Curve Graph | Display |
| Printer | optional build-in thermal printer |
| COM | USB |
| Software | Optional software connecting with computer |
| Power supply | AC 220 V± 22 V 50/60 Hz ± 1 Hz, 250 W |
| Packing Dimensions | 780 × 560 × 390 mm |
| Gross weight | 18 Kg |
Determines concentration of metals like K, Na, Li, Ca, Ba
Uptakes sample up to < 6 L/min
Calculates correlation coefficient automatically
Concentration units are selectable
Measuring range can be altered
Flame size can be pre-selected
7-inch color touch screen to display concentration values
Multilingual interface for users convenience
Flame Photometer LBN-FP170 is widely used for quantitative analysis of alkali and alkaline earth metals across pharmaceutical industry, food and beverage industry, petrochemical industry and in research fields.