Variable Volume Pipette LBN-VP102 can dispense liquid volumes within the range of 2 to 20 μL ensuring accuracy and consistency. It is crafted to withstand autoclaving at 121℃, ensuring complete sterilization. The tip cones are made from chemical-resistant materials and employ a spring-loaded design to facilitate easy maintenance. Our pipette offers easy calibration and maintenance, providing precise results and user-friendly operation.
Volume Range | 2 to 20 μL |
Increment | 0.5 μL |
Test Volume | 20 μL, 10 μL, 2 μL |
Accuracy Error | 20 μL: 0.18 μL, 0.90% 10 μL: 0.12 μL, 1.20% 2 μL: 0.06 μL, 3.00% |
Precision Error | 20 μL: 0.08 μL, 0.40% 10 μL: 0.1 μL, 1.00% 2 μL: 0.04 μL, 2.00% |
Equipped with a digital display
Incorporates a low-force tip ejector
High-sealing performance of tip cones
Rotating dispensing head for easy pipetting
Provides finger support for user comfort
Adheres to ISO 8655 standards
Variable Volume Pipette is employed for precise measurement and transfer of liquid volumes. It is used in various scientific fields such as biology, chemistry, biochemistry, and molecular biology.
Download Catalog for Variable Volume Pipette LBN-VP102
DownloadVolume Range : 10 to 100 μL
Increment : 1 μL
Test Volume : 100 μL, 50 μL, 10 μL
Accuracy Error : 100 μL: 0.8 μL, 0.80%
50 μL: 0.5 μL, 1.00%
10 μL: 0.3 μL, 3.00%
Volume Range : 0.5 to 10 μL
Increment : 0.1 μL
Test Volume : 10 μL, 5 μL, 1 μL
Accuracy Error : 10 μL: 0.1 μL, 1.00%
5 μL: 0.075 μL, 1.50%
1 μL: 0.025 μL, 2.50%
Volume Range : 5 to 50 μL
Increment : 0.5 μL
Test Volume : 50 μL, 25 μL, 5 μL
Accuracy Error : 50 μL: 0.3 μL, 0.60%
25 μL: 0.225 μL, 0.90%
5 μL: 0.1 μL, 2.00%
Volume Range : 20 to 200 μL
Increment : 1 μL
Test Volume : 200 μL, 100 μL, 20 μL
Accuracy Error : 200 μL: 1.2 μL, 0.60%
100 μL: 0.8 μL, 0.80%
20 μL: 0.6 μL, 3.00%
Volume Range : 50 to 200 μL
Increment : 1 μL
Test Volume : 200 μL, 100 μL, 50 μL
Accuracy Error : 200 μL: 1.2 μL, 0.60%
100 μL: 0.8 μL, 0.80%
50 μL: 0.5 μL, 1.00%
Volume Range : 200 to 1000 μL
Increment : 5 μL
Test Volume : 1000 μL, 500 μL, 200 μL
Accuracy Error : 1000 μL: 6 μL, 0.60%
500 μL: 3.5 μL, 0.70%
200 μL: 1.8 μL, 0.90%
Volume Range : 1000 to 5000 μL
Increment : 50 μL
Test Volume : 5000 μL, 2500 μL, 1000 μL
Accuracy Error : 5000 μL: 25 μL, 0.50%
2500 μL: 15 μL, 0.60%
1000 μL: 7 μL, 0.70%
Volume Range : 2 to 10 mL
Increment : 0.1 mL
Test Volume : 10 mL, 5 mL, 2 mL
Accuracy Error : 10 mL: 60 μL, 0.60%
5 mL: 60 μL, 1.20%
2 mL: 60 μL, 3.00%
Pipette Tip Volume : 1000 µL
Pack Size : 100 holes
Pieces : 100pcs
Color : Blue